![Froilabo BM340 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/froilabo/bm340/bm340_user-manual_2338685014.webp)
10
Maximum power consumption
110V (Watts) (BoSS System)
1600
1700
2400
1700
2400
1700
Power consumption* 110V, 60
Hz (Watts)
1000
1100
1400
1100
1400
1100
Minimum power of cooling of
the room containing the freezer
in continuously (W)
850
950
1200
950
1200
950
Acoustic pressure measured in
dB(A) (BoSS System)
76
Acoustic pressure measured in
dB(A) (in regulation mode)
56
OUTSIDE DIMENSIONS
Height (mm)
1280
1280
1640
1640
2000
2000
Width (mm)
875
875
875
875
875
875
Depth (mm)
970
970
970
970
970
970
Clearance required at the rear of
the unit (mm)
200
200
200
200
200
200
Maximum dimension with door
open (mm)
1750
1750
1750
1750
1750
1750
Maximum width with door open
(mm)
1150
1150
1150
1150
1150
1150
Maximum door opening angle
110°
110°
110°
110°
110°
110°
GROSS INTERNAL
DIMENSIONS EXCLUDING
RACKS
Height (mm)
716
716
1076
1076
1436
1436
Width (mm)
630
630
630
630
630
630
Depth (mm)
752
752
752
752
752
752
Notes
* Power consumed in regulation mode, all parts activated and stabilized, at 23 ° C in atmosphere and with empty cabinet.
On an equivalent product, hydrocarbon (HC) gases allow a reduction in power consumption up to 25%, as the need of cooling.
Refrigeration system:
FREEZER MODEL
BM340
BM515
BM690
Standard
cooling
Natural
cooling
Standard
cooling
Natural
cooling
Standard
cooling
Natural
cooling
TWO-STAGE -86°C MODELS
Stage 1 hermetic compressor
power (Watt)
580
380
780
580
780
580
Stage 2 hermetic compressor
power (Watt)
380
380
580
580
580
580