20
Display of radiation risk warning: ON/OFF.
Default value: OFF.
When set to ON, the logo is displayed at the top of the screen.
Return to previous menu
7.5
« S
AMPLE LOCATION
»
MENU
The main menu gives access to the three functions of samples location. This menu is an option. Once the option
is activated, from the “Home screen”, access the “Location sample menu” by pressing on the following
pictogram.
Location sample menu
Search for a sample (protected by a password)
Enter a new sample
Configure the freezer storage space
Return to the previous menu without saving changes
Pressing one of these buttons brings up a username entry screen
a)
Configuring the freezer :
The freezer must be configured to specify the type of storage elements incorporated. Configuration therefore
entails defining the types and locations of drawers and shelves.
From the “Location sample” screen, access the freezer configuration menu by pressing on the following
pictogram (on the right).
Enter the password and validate.
Select the size of the freezer by choosing the
appropriate model. The “elementary” grid
corresponding to the capacity is displayed
(factory-set default value).
Note: several sample numbers can be stored in the
same location.
Select a drawer two units wide.
Select a rack one unit wide.
Select any other accessory two units wide.