Actions
Alarm condition
Pumping
status
Symbol
■
Stop infusion or restart a new
one.
The target volume is
reached. (Complete
progress bar)
Pumping stops
if KVO
deactivated or
pumping
continues with
KVO flow rate
End of infusion
■
Connect the pump to the
mains via the pump holder.
Charge battery to resume
pump operation.
Minimum battery voltage
is not available. Appears
approx. 10 min. before
battery is fully discharged.
Pumping stops
Empty battery
Low priority - Yellow LEDs are fixed and alarm sound (sequences of 3 beeps)
■
Get prepared for end of
infusion
Target volume will be
reached.
The time of message
before target volume is
reached can be set in the
menu.
See
on page 63.
Pumping
continues
End of infusion
pre-alarm
83