Programming screen / Configuring
screen
2nd level menu
1st level menu
Select occlusion pressure level
between high and low
Set single bubble volume detection
(range: 0.125 mL to 0.75 mL)
Set cumulated bubbles volume
detection (volume range: 0.25 mL to
0.75 mL) / (duration range: 1 min to
30 min)
Occlusion pressure / Air detection
Advanced
settings menu
(access code
required)
Activate / Deactivate end of infusion
pre-alarm
Select end of infusion pre-alarm
mode setting between remaining time
and remaining volume (when the end
of infusion pre-alarm is actived)
Program the value according to the
selected mode
End of infusion pre-alarm
Activate / Deactivate nurse settings
lock
Nurse settings lock
Activate / Deactivate KVO
Adjust KVO rate value (range: 1 mL/h
to 10 mL/h)
KVO (keep vein open)
Set maximum flow rate (range: 1
mL/h to 600 mL/h)
Maximum flow rate
Reset pump to factory settings
Reset manufacturing settings
-
Infusion history (consult the last 250 infusion
events)
History menu
-
Cumulative counter (cumulated counter from
1 day to 99 days)
-
Alarm history (consult the last 250 infusion
events)
42