2.
During cleaning, do not turn the pump upside-down to avoid liquid leak into the pump.
3.
Use a ready-to-use wipe to remove any major grime.
4.
Thoroughly wipe down all exposed surfaces (housing, keypad, screw area, holder
connection area, etc.) of the pump, from top to bottom. Gently wipe down the pump
exposed mechanism and sensor area (tube guide, blue insert).
A minimum cleaning of 1 minute is recommended (allow to remain visibly wet for 1
minute), until all organic matter is dissolved and removed. Do not allow liquids to run,
leak, or drip into the pump housing. Use cotton wool to clean the contact pins.
5.
Repeat step 4 with the pump door and holder.
The door can be removed from the pump to facilitate the cleaning.
NOTE:
The door can be immersed. Clean it separately with running water.
6.
Using a fresh ready-to-use wipe, thoroughly wipe down all exposed surfaces. A minimum
cleaning of 1 minute is recommended (allow to remain visibly wet for 1 minute), until all
organic matter is dissolved and removed.
7.
Wipe down the power cable.
8.
Allow the device to dry completely at room temperature.
NOTE:
After step 8, perform disinfection procedure described in
on page 75.
7.4 Disinfection instructions
Prerequisites
■
The cleaning protocol has been performed.
■
The pump is disconnected from the patient.
■
The pump is switched off.
■
The power cable and all other cables are unplugged.
■
The pump is disconnected from the holder.
75