Actions
Alarm condition
Pumping
status
Symbol
■
Open the door, check that the
set is installed properly, close
the door.
See
on page 27.
■
Check that the set is not
kinked.
■
Check that the infusion tube is
clear.
■
Flush tube if necessary.
■
Re-position and verify that fluid
flows freely after adjustment.
■
Check the absence of
upstream / downstream
occlusion in the line.
Downstream flowpath is
blocked after the pump, at
the patient side.
Pumping stops
Downstream
occlusion
■
End infusion or connect to a
filled container.
Infusion container is
empty.
Pumping stops
Empty bag / Air in
line
and
appears
alternatively
■
Fill set to the end.
See
on page 29.
Air is in the set.
■
Open the door and remove dirt
with cloth and soapy water or
as directed by healthcare
facility policy or healthcare
provider policy (see
on page 74).
■
Allow the pump to dry.
Dirt in sensor area (lower
tube guide).
■
Check position of the set and
insert correctly if necessary.
See
on page 27.
The set not properly
connected to the
container.
82