ERROR MESSAGES
MESSAGE
PROBABLE CAUSE
SOLUTION
Error - 000
QC has to be done every
12 or 24 hours. If QC
is overdue, “Error” will
appear. After 3 seconds,
device will enter QC Mode.
Please proceed with QC Mode.
Error - 002 Battery voltage too low
Please recharge the battery.
Error - 004 Operator ID error
Please scan or enter operator
ID again.
Error - 005 Patient ID error
Please scan patient's ID again.
Error - 006 Test strip error
Please scan test strip ID again.
Error - 007 Test strip has been used
Please insert a new test strip
and repeat test.
Error - 008 Test strip was pulled out
Please re-insert the test strip.
Error - 009 Low level control solution
error
Please scan low level control
solution lot number again.
Error - 010 Control solution expired
Please use a new vial of control
solution and repeat test.
Error - 011 Test strip lot error
Please check strip lot number
and repeat the test.
Error - 012 Test strip expired
Please use a new strip and
repeat test.
Error - 013 QC result out of range
Please check control solution
vial label and repeat the test.
Error - 018 Ambient temperature is too
low.
Please operate the meter under
the temperature of 50°F - 104°F
(10°C - 40°C).
Error - 019 Ambient temperature is too
high.