37
WARNING: Potential Biohazard!
When to clean and disinfect the meter:
How to clean and disinfect the meter:
Healthcare professionals using this system on multiple patients
should follow the infection control procedure approved by their
facility. All products or objects which come in contact with human
blood, even after cleaning, should be handled as if capable of
transmitting viral diseases.
To avoid the meter and test strips attracting dirt, dust or other
contaminants, please wash hands thoroughly with soap and water
before and after use.
All surface of meter if visibly soiled must be physically cleaned to
remove gross soil. Disinfect the meter between each patient to
prevent infection.
The meter must be cleaned prior to the disinfection. Use one
disinfecting wipe to clean exposed surfaces of the meter thoroughly
and remove any visible dirt, blood, or any other body fluid with
the wipe. Use a second wipe to disinfect the meter by following
the disinfecting procedure below. Do NOT use organic solvents to
clean the meter.
We recommend for meter cleaning and disinfection you should
use the disinfecting wipes/towelettes from below which the active
ingredients have been tested to be effective against Hepatitis B
Virus (HBV) for
ForaCare
POCT S10 blood glucose meters:
Micro-Kill+™ (Micro-Kill Plus™) by Medline (EPA Reg. No.
59894-10-37549)
To obtain disinfecting wipes and other information, please
contact Medline at 1-800-MEDLINE (1-800-633-5463) or visit
www.medline.com.
The Micro-Kill Plus disinfectant product has been shown to be safe for use with
ForaCare
POCT S10 meter, but any disinfectant product with the EPA registration number of (EPA
Reg. No.59894-10-37549) may be used on this device.
TAKING CARE OF THE METER & STRIP