FONA XPan 3D Operating Instructions
24/28
69 683 70110
-
150831
9.
Program Values
Factory programmed values
Freely programmed values
Program
P1, P3, P4, P5, P6, P7
61/6.3 67/6.3 73/6.3 79/6.3
P2
61/5.0 67/5.0 73/5.0 79/5.0
P8, P9, P10
61/6.3 67/6.3 73/6.3 79/6.3
10. Care of the surfaces
ATTENTION
Always disconnect the system from the mains (main switch in the
room) before cleaning it.
Cleaning
Use a mild soap to remove fingerprints or other traces of dirt being
careful not to let liquid substances penetrate the machine.
The plastic covers can be cleaned with a soft cloth and a mild
detergent.
Disinfecting
The parts that come into contact with the patient must be cleaned
after each use with a detergent (for example, a 2% ammonia
solution) and then disinfected. DO NOT use solvents or corrosive
substances.
The bite block and the chin rest can be sterilized in autoclave at
134°C.
11. Inspection and maintenance
Inspection and maintenance work must be performed at regular
intervals to protect the safety and health of patients, users and third
parties.
Annual inspection performed by the
operator or other authorized
personnel
As the operator, you should ensure the safety and reliability of your
system by performing maintenance on it at regular intervals (at least
once annually) or having this work performed by your dental
dealership.
Maintenance performed by the
service technician
In addition to the scheduled annual inspection by the user or persons
contracted to perform this, a maintenance inspection must be
performed by the service technician after 4, 7, 10 years and then
every two years.
Checking image quality
At regular intervals, however at least once a year, the user must
evaluate the image quality.