ATTENTION. Do not switch off the unit or the computer to
prevent the saved image to be deleted, thus making re-transmission
possible.
Exposure interruption
ATTENTION. For safety reasons the operator can terminate the
exposure any time by releasing the exposure switch.
Premature termination is signaled by an error message:
Error 11: Exposure aborted during irradiation.
The patient has been partially exposed. Part of the radiograph is being
available. Examination may have to be repeated.
Error 12: Exposure aborted before irradiation.
The patient was not irradiated.
Examination to be restarted.
Error 20: Exposure aborted after irradiation.
No need to repeat the examination.
Data Acquisition System Not Ready.
The data acquisition program has to be started first to make sure data
acquisition will be performed during radiographic exposure.
Automatic Exposure Blockage
ALARM RED LIGHT BLINKING
This feature protects the X-ray tube by preventing premature
triggering of a new exposure should the load requested by the next
exposure exceed the available capacity.
The red light on the control panel keeps blinking until cool-off time
has elapsed.
Other error messages Possible malfunctions during the use generate an error message and
the unit is blocked. A list of messages are reported in chapter 12.
Depending on the malfunction, technical service might be required.
7.5 Remote hand-switch
The hand switch can alternatively be mounted remotely in case the
unit is located in an X-ray room which has a door and enables visual
contact with the patient.
An optional kit is available for remote mounting of the hand switch.
Order code 93 190 00155.
Wait PC READY
ERROR
#: 20
ERROR
#: 12
ERROR
#: 11