FONA XPan 3D Operating Instructions
69 683 70110
- 150831
17/28
D.
Examination Selection
Key for PROGRAM selection,
to sequentially change the program from 1 to 10 and back again.
Key for PATIENT build selection,
Small, Medium, Large, Extra Large.
The pre-programmed technique factors are selected.
Manual correction of tube voltage and of tube current can be done using the
INCREASE or DECRESE keys at display sides (or PC side).
READY GREEN LIGHT
ON
RETURN Arm Movement to bring the arm from PATIENT ENTRY position to START
position, ready to start the exposure. When the unit reaches the START position the
green light of the READY indicator on the control panel is turned ON.
TEST Mode without Radiation
With unit in START position and the READY light ON, TEST mode can be activated to
run the unit without radiation.
On the display are shown:
The selected program number
The message “Test Mode” instead of values of kV and mA
A green light is also activated, above the key button
Starting the unit with the hand-switch allows for rotation of the arm according to the
program selected.
When the arm is returned to PATIENT ENTRY position test mode is terminated and
the units enters normal mode.
E.
Setting Tube Voltage
Key INCREASE at the left of the display
to manually raise kV level.
Key DECREASE at the left of the display
to manually lower kV level.
The tube voltage can be set from 61 to 85 kV in steps of 3 kV.
61
64
67
70
73
76
79
82
85
F.
Setting Tube Current
Key INCREASE at the right of the display
to manually raise mA level.
Key DECREASE at the right of the display
to manually lower mA level.
The tube current can be set from 4 to 10 mA.
4.0
5.0
6.3
8.0
10
P1
-- TEST MODE --