FONA XPan 3D Operating Instructions
69 683 70110
- 150831
23/28
8.
Programming
Programming procedure
In the factory kV and mA values are assigned to ADULT panoramic
programs (P1, P3, P4, P5, P6, P7) and CHILD panoramic program (P2)
for each one of the four types of patient build, for a total of 40
combinations.
Such values can be re-loaded using a dedicated service function
Entering Service Mode
By keeping pressed the BACK key for 5 s the system enters Service
Mode
Service function can be selected using
the INCREASE or DECRESE keys on the RIGHT side of the display.
Scroll the menu until Exposure Settings is displayed
Press the PROGRAM key to enter Exposure Settings.
NOTE. Keep pressed the TEST key to reset all values to default
(factory) conditions
By pressing the PROGRAM key select the exposure program
for ADULT panorama (P1, P3, P4, P5, P6, P7) or
for CHILD panorama (P2) or for 3D exam (P8, P9, P10).
By pressing the PATIENT key select the patient build,
Small, Medium, Large, Extra Large
Change the kV level to the desired value by pressing the
INCREASE or DECRESE keys at the LEFT side of the display
Change the mA level to the desired value by pressing the
INCREASE or DECRESE keys at the RIGHT side of the display
Keep pressed the LIGHT
key to store the selected value
Repeat kV setting and mA setting
for the other types of patient build if requested and for the other
CHILD or ADULT group of programs if requested
Press BACK key to exit EXPOSURE SETTINGS
Press TEST Mode key to exit SERVICE MODE
Do verify proper setting by changing PATIENT build for ADULT
panorama (programs P1, P3, P4, P5, P6, P7), CHILD panorama
(program P2) and 3D (P8, P9, P10)
* SET kV / mA *
67 6.3
S/-
Expos. Settings