FONA CDRelite System User Guide
B1057003 Rev. 3
31
6 .
If the Sensor or cable are visibly soiled (e.g., with blood or
saliva), each should be cleaned with a soapy cloth or paper
towel, and then dried with a clean lint-free cloth or paper
towel.
7 .
Thoroughly wipe the Sensor and cable with the disinfecting
product recommended in Section 6.3. Do not expose the
Sensor cable connectors to any amount of liquid.
8 .
Allow the liquid disinfectant to remain on the Sensor for 30
seconds.
9 .
Repeat steps 7 and 8. When the Sensor has been wiped two
times, continue with the following steps.
10 . Remove potential chemical build-up from the Sensor by
wiping it with a sterile lap sponge saturated with de-ionized
water.
11 . Use a sterile dry lap sponge to dry the Sensor or cable, as
needed.
12 . Place the Sensor in a clean environment, ready for next use.
13 . Reconnect the Sensor.
14 . Remove and discard gloves.
6.3. Recommended Disinfectants
The following surface disinfectant has been found to be effective in
achieving a desired level of disinfection and is available from other
dental product suppliers.
Cavi-Wipes (Metrex Research, Kerr) or equivalent