IMPORTANT! To avoid exposing the patient to unnecessary
X-rays, ensure that the CDR exam window viewbox is flashing
green (default color) in AutoTake mode, or that the message,
“Waiting to take X-ray,” is displayed before triggering the X-ray
Source.
A.
If your acquisition mode is set to AutoTake, the first empty
view box in the exam is pre-selected and flashes green
(default setting). Skip ahead to step 7.
B.
If your acquisition mode is set to manual, select an empty
target frame that corresponds to the Sensor’s location in the
patient’s mouth. When the view box is highlighted, click on
it again. If using a serial footpedal, press the amber pedal.
S
TEP
6
The system is ready to acquire an X-ray image. A “Please
Wait” message may appear momentarily.
S
TEP
7
A.
In AutoTake mode, activate the X-ray source. The
message, “Reading Image from Sensor” appears
momentarily.
B.
In manual mode, activate the X-ray source when the
message, “Waiting to take X-ray” appears.
C.
The acquired image appears automatically in the zoom or
exam window, depending on system settings.
S
TEP
8
A.
In AutoTake mode, the next empty view box in the series
sequence is selected. To acquire the next X-ray image,
repeat this procedure starting at Step 7.
B.
In manual mode, close the exam window. To acquire the
next X-ray image, repeat this procedure starting at Step 5.