8. Cable Replacement
IMPORTANT!
Always disconnect the Sensor from the FONA CDRelite USB interface
during cable replacement to avoid potential damage to Sensor components. Close CDR
DICOM (or other imaging software) prior to starting cable replacement. When performing
cable replacement, always work outside the patient area, using the tools and materials
supplied and / or recommended by Sirona Dental.
IMPORTANT!
Like other electronic devices, your FONA CDRelite Sensor is susceptible to
electrostatic discharge (ESD), particularly when the device is used in or around carpeted
areas or low-humidity environments. During cable replacement, when Sensor contacts are
exposed, it is especially important to protect the device from potential ESD damage.
Touching a metal surface prior to replacing the FONA CDRelite cable will reduce the risk of
damaging Sensor components by accidental static discharge. The use of anti-static floor
mats or floor treatments (for example Staticide 2005/2002) will also help eliminate static
build-up in your office.
S
TEP
1
A.
Please clean and dry your hands before
performing this procedure. Do not wear
gloves since the powder inside them
could be deposited on sensor contacts
while replacing the cable.
B.
Make sure the Sensor is placed securely
on a clean, moisture-free surface.
C.
Using a dental instrument, carefully lift
and remove the tab cover from the back
of the Sensor cable. Dispose of the tab
cover as a new one will be used when
the new cable is attached.
S
TEP
2
A.
Using the screwdriver provided, loosen
and remove the 2 screws that secure the
cable to the Sensor. Dispose of the
screws as new ones are supplied.
B.
Remove the cable from the Sensor.