Operating Manual EMGZ622A/642A
22
9.5
Description of the Parameters EMGZ622A/642A
The parameter changing mode is activated by pressing the
PARA
↵
key for 3 seconds. The
required module is then searched with the
↑
↓
keys and selected with the
PARA
↵
key (ref.
also to fig. 12).
Offset
Use:
This parameter stores the value determined with special function
Find
offset
in [Digit]. It is not necessary to note this parameter because a new
offset adjustment is done very easy; also when changing the whole
electronic unit.
The offset can also be inputted manually with the
↑
↓
←
keys.
Range:
–8000
to
8000
Default:
0
Increment:
1
Unit:
[Digit]
Gain
Use:
This parameter stores the value determined with special function
Calibration
, or you must input a value calculated using the formulas
written under „8.3 Calibrating the measuring amplifier“ if the
material
tension
cannot be simulated.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Gain 1
Use:
Identical
with
Gain
but the value stored here was determined by special
function
Calibration gain 1
. The value stored here is used if the digital
input „switch gain“ is activated (ref. to wiring diagram).
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Force of sensor
Use:
This parameter stores the nominal force of the sensor. It is printed to the
nameplate of the force sensor.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N, kN, cN]
Unit of sensor
Use:
This parameter stores the measuring unit of the sensor. It is printed to the
nameplate of the force sensor.
Range:
N, kN, cN
Default:
N
Содержание EMGZ622A
Страница 8: ...Operating Manual EMGZ622A 642A 8 6 Dimensions fig 3 Dimensions E611002e...
Страница 26: ...Operating Manual EMGZ622A 642A 26 10 Serial Interface RS232 Optional...
Страница 28: ...Operating Manual EMGZ622A 642A 28...
Страница 31: ...Operating Manual EMGZ622A 642A 31 13 Interface CAN Bus Optional...
Страница 32: ...Operating Manual EMGZ622A 642A 32 14 Interface DeviceNet Optional...