Operating Manual EMGZ622A/642A
14
8.2
Configuring the Electronic Unit
The use of the input channels provided is as follows:
Input Channel assignment (ref. also to wiring diagram)
EMGZ622A
EMGZ642A
Channel 1
Range 1A (low force range)
Range 1A (measuring point 1, low
force range)
Channel 2
Range 1B (high force range)
Range 1B (measuring point 1, high
force range)
Channel 3
–
Range 2A (measuring point 2, low
force range)
Channel 4
–
Range 2B (measuring point 2, high
force range)
Prior to the first calibration, the following settings must be done for each channel (ref. to „9.
Parametrization“):
Global parameters
Language
Required display language
Measuring system
Metric (default) or US standard
Channel parameters
Force of sensor
Ref. to nameplate of the force sensor
Unit of sensor
Ref. to nameplate of the force sensor
Sensitivity
FMS force sensors = 1.8mV/V (default)
Scale output
Which tension feedback refers to 10V resp. 20mA?
Config. output
0...10V (Default) or ±10V,
corresponding to parameter
Config. output
Note
Wrong setting of the parameters may cause malfunction of the electronic unit! Setting of the
parameters must be done carefully prior to setting into operation!
Содержание EMGZ622A
Страница 8: ...Operating Manual EMGZ622A 642A 8 6 Dimensions fig 3 Dimensions E611002e...
Страница 26: ...Operating Manual EMGZ622A 642A 26 10 Serial Interface RS232 Optional...
Страница 28: ...Operating Manual EMGZ622A 642A 28...
Страница 31: ...Operating Manual EMGZ622A 642A 31 13 Interface CAN Bus Optional...
Страница 32: ...Operating Manual EMGZ622A 642A 32 14 Interface DeviceNet Optional...