Operating Instructions EMGZ492.PNET
05.10.2018
27
0x0D
Limit frequency low-pass filter actual value B
Limit frequency of the low-pass filter for the actual value
outputted via PROFINET.
Access
type
R/W
Parameter command
limit frequency low-pass filter
actual
value
(PROFINET)
Data type
int (unsigned 16 bit)
Value range
1 to 2‘000
Value format
###.#
Unit
Hz
0x0E
Offset adjustment B
Determine and store offset. The system is set to zero
without material tension.
Access
type
W
Parameter command
offset adjustment
Data type
byte (unsigned 8 bit)
Value range
0 to 1
Value format
#
0x0F Calibration
B
Calibrates the amplifier to the weight in Newton, which is
handed over here. It must match the suspended weight.
Access
type
W
Parameter command
calibration
Data type
long (signed 32 bit)
Value
range
±
200‘000‘000
Value format
######.###
Unit
N
Explanation of access types: R = Read, W = Write, R/W = Read and Write.
To address the parameters 0x20 to 0x23 you have to use slot 0, module <device
name>, “Parameter Access Point”.