Operating Manual EMGZ309._.EIP
10
4
Configuring the Amplifier
The amplifier EMGZ309._.EIP can be configured over the operation panel (chapter
8), over a PC (web interface, chapter 9) or over EtherNet/IP. The basic configuration
is performed over the operation panel.
4.1
Power up the EMGZ 309._.EIP
1.
Connect the first force sensor (see Fig. 4 Wiring Diagram EMGZ309)
2.
Check whether applying a force in measuring direction (in the direction of the red
point) on the first sensor results in a positive output signal. If not, exchange the
two signal wires of the corresponding force sensor (terminals 12/13).
3.
If applicable, connect the second force sensor.
4.
Check whether applying a force in measuring direction on the second sensor
results in a positive output signal. If not, exchange the two signal wires of the
corresponding force sensor in the terminal block (terminals 22/23)
4.2
Operating Panel
Fig. 5: Operating Panel EMGZ309.W.EIP
E309007e
The keys on the operating panel have the following functions:
Key
Key Function
Change the selections or increase / decrease the values
Enter
>0< + PARA
Enter Offset Compensation Procedure. Press the two keys at
the same time for longer then 3 sec.
Cal + PARA
Enter Calibration Procedure. Press the two keys at the same
time for longer then 3 sec.
PARA
Enter the Parameter Setting Mode. Press the key at the same
time for longer then 3 sec.