Operating Manual DLS2
19
9.2
Call Profile and Implement
If a profile is fetched from the memory and adopted, the actual calibration data of the
DLS2 is overwritten.
1.
Enter operation mode „Profile Management“ by pressing the
key for longer
than 0.5 sec.
2.
Select an existing profile with the
keys. An existing profile is marked by
shining LED.
3.
Press the
key for longer than 3 sec. to adopt the selected profile. This action
terminates the Profile Management mode and brings the DLS2 to normal operation.
9.3
Save Profile
1.
Calibrate DLS2 with material-reference-combination.
2.
Check calibration with a test run.
3.
Enter operation mode „Profile Management“ by pressing the
key for longer
than 0.5 sec
.
4.
Select a free profile storing place with the
keys. A free profile storing place is
marked by a flashing LED. It’s not possible to overwrite an existing profile without
deleting the profile beforehand. .
5.
Press the
key for longer than 0.5 sec. to save the profile in the free storing
place. The LED will change from flashing to permanent shining. If the saving
process was successful, the DLS2 will go to normal operation mode.
9.4
Delete Profile
1.
Enter operation mode „Profile Management“ by pressing the
key for longer
than 0.5 sec
.
2.
Select an existing profile with the
keys. An existing profile is marked by
shining LED.
3.
Press the
key for longer than 3 sec. to delete the selected profile. The LED
will change from permanent shining to flashing. The sensor will then go to normal
operation mode.