Operating Manual BMGZ710 and BMGZ710.PNET
20.04.2021
22
start zero settubg or adjust offset
manually.
4 Impulse output too fast.
The impulse output is no longer correct. The
load is higher that the output can indicate.
Reduce speed or load or enhance the
parameter of the impulse output [kg] so
that more weight per pulse can be
outputted.
Device offset
In the event of exchange or repairs, this can
only be changed via web browser with a
password.
Device gain
In the event of exchange or repairs, this can
only be changed via web browser with a
password.
Totalizer
In the event of exchange or repairs, this can
only be changed via web browser with a
password.
6.5
Digital inputs
The functions of the digital inputs are preconfigured and cannot be changed.
6.5.1
Digital input 1 (zero setting)
Setting this input begins the procedure for zero setting. Another activation aborts the
ongoing procedure.
6.5.2
Digital input 2 (batch active)
Setting this input starts the weighing of a new batch. The batch counter is incremented,
and the flow rate is set to zero.
If the digital input 'Batch active' is set, the batch counter (quantity) is zeroed (identical
to when the Reset softkey is pressed) and the batch no. is increased by 1. While this
input is active, a batch is weighed. If the input becomes passive, the batch no., the
weighed quantity (batch counter), date and time are saved in the alibi protocol. This
ensures that weighed batches are traceable.
6.5.3
Digital input 3 (belt running)
The “belt running” signal tells the evaluation unit that the flow rate must be integrated
over time.
6.5.4
Digital input 4 (impulse)
4 "impulse” signals tell the evaluation unit that the measuring roller has completed one
revolution.
If the v-detection parameter is set to 'None', the evaluation electronics does not
calculate the speed using the initiator information, but takes 1.00 m/s as the belt