Operating manual BMGZ 600A
38
The service mode 1 contains parameter for checking the AD value of the force
sensors which are integrated into the measuring roller. The service mode 1 is
activated by pressing the CLR and PAR keys for 3 seconds.
AD value roller 1/2
Use:
This parameter shows the source signal of the corresponding
measuring roller before offset compensation. It may be helpful for
trouble shooting.
Range:
–8192
to
8191
Unit:
[Digit]
(to be inspected only)
9.6
Service Mode 2
fig. 27: Service mode menu 2 BMGZ610A/611A. The versions BMGZ620A have
its specific but equivalent service mode menu.
B600027e
The service mode 2 contain the parameter to set the totalizator, when changing
the electronic unit for example. The service mode should only be activated by
trained service personal. The service mode 2 is activated by pressing the
0
and PAR keys for 3 seconds. Generally, the service parameter can be changed the
same way as the other parameters.
Set Totalizator 1/2
Use:
When replacing an evaluation electronic unit, the values of the
totalizators can be taken from the old to the new electronic unit.
Therefore, note the totalizator values before dismounting the old
electronic unit and store the values here after mounting the new
unit.
Range:
0
to
1000000
Default:
0
Increment:
1
Unit:
[t]
Содержание BMGZ 610A
Страница 6: ...Operating manual BMGZ 600A 6 components in italic letters are option...
Страница 16: ...Operating manual BMGZ 600A 16 Repeat the procedure for the 2nd side disk...
Страница 18: ...Operating manual BMGZ 600A 18 8 Replace the earth cable and all the cables with connectors back to place...
Страница 20: ...Operating manual BMGZ 600A 20 6 6 Wiring Information fig 16 Terminal arrangement on terminal board B600007...
Страница 45: ...Operating manual BMGZ 600A 45 fig 31 Flat measuring roller B400013e...