![Fluke TS44DLX Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/fluke/ts44dlx/ts44dlx_user-manual_2310025021.webp)
17
Maintenance
Removing the Old Line Cord
To remove the old line cord:
W
Caution
Be sure to account for all hardware
removed. Loose or missing hardware could
create a hazard for the end user.
Notes
Battery cover screws are retained in the cover
and will not come all the way out of the battery
cover.
Be careful not to damage or pinch the speaker
wires, printed circuit faces or insulating
materials.
1
Use a Phillips screwdriver to loosen only the four
screws attaching the battery door cover to the
back of the test set (See Figure 4).
2
Remove the battery door cover (see Figure 4) and
battery from the test set.
3
Loosen the two screws that hold the line cord to
the PCB (see Figure 6).
4
Using needle nose pliers or one of the line cord
clips, remove the two screws and washers from the
line cord connectors.
5
Using needle nose pliers, remove the plastic clip
(see Figure 7) that holds the line cord strain relief
in place.
6
Slip the line cord screw lugs (see Figure 8) out
through the hole in the end of the housing.
bfp06.eps
Figure 6. Removal/Installation of Line
Cord Screws
bfp07.eps
Figure 7. Removal/Installation of Plastic Clip
9 VOL
T