T6-1000/T6-600
Service Information
4
Specifications
Battery Type and Life
............................................2 x AA (IEC LR6)
360 hr continuous, typical is 200 hr using the FieldSense function
Fork Opening
.........................................................17.8 mm
Temperature
Operating .............................................................-10 °C to +50 °C (+14 °F to +122 °F)
Storage ................................................................-30 °C to +60 °C (-22 °F to +140 °F)
Operating Relative Humidity
0 % to 95 %..........................................................5 °C to 30 °C (41 °F to 86 °F)
0 % to 75 %..........................................................30 °C to 40 °C (86 °F to 104 °F)
0 % to 45 %..........................................................40 °C to 50 °C (104 °F to 122 °F)
Altitude
Operating .............................................................2000 m
Storage ................................................................10 000 m
Temperature Coefficient
.......................................0.1 x (specified accuracy) / °C for <18°C or >28 °C (<64.4 °F or >82.4 °F)
Safety
General ................................................................IEC 61010-1: Pollution Degree 2
Measurement .......................................................IEC 61010-2-032, IEC 61010-2-033
T6-600 ..............................................................CAT III 600 V
T6-1000 ............................................................CAT IV 600 V / CAT III 1000 V
Electromagnetic Compatibility (EMC)
International .........................................................IEC 61326-1: Portable Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio
frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those
directly connected to a low-voltage power supply network that supplies buildings used for
domestic purposes. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.
Korea (KCC)........................................................Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment
and the seller or user should take notice of it. This equipment is intended for use in
business environments and not to be used in homes.
USA (FCC) ..........................................................47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Model T6-1000
Model T6-600
Function
Requires Test
Leads
Range
Range
Resolution
Accuracy
[1]
FieldSense
ac voltage true-rms
No
1000 V
600 V
1 V
±(3 % + 3 counts)
45 Hz to 66 Hz
[2][3]
FieldSense
ac current true-rms
No
200.0 A
200.0 A
0.1 A
±(3 % + 3 counts)
45 Hz to 66 Hz
FieldSense
Frequency (Hz)
No
45 Hz to 66 Hz
1 Hz
±(1 % + 2 counts)
[3]
Volts ac true-rms
Yes
1000 V
600 V
1 V
±(1.5 % + 2 counts)
45 Hz to 66 Hz
Volts dc
Yes
1000 V
600 V
1 V
±(1 % + 2 counts)
Resistance
Yes
2000
Ω
2000
Ω
1
Ω
±(1 % + 2 counts)
Yes
20.00 k
Ω
0.01 k
Ω
Yes
100.0 k
Ω
0.1 k
Ω
[1] Accuracy: ±( [ % of reading] + [number of least significant digits] ). Accuracy is specified for 1 year after calibration, at 18 °C to 28 °C
(64 °F to 82 °F) with relative humidity to 90 %. AC measurements are ac coupled, RMS responding.
[2] Add 3 % typical without an external ground connection. External ground connection required for user wearing insulated gloves,
standing on an insulated ladder, or otherwise insulated from earth ground.
[3] FieldSense is specified from 16 V to 100 % of range.