background image

T6-1000/T6-600

Service Information

4

Specifications

Battery Type and Life

............................................2 x AA (IEC LR6)

 

360 hr continuous, typical is 200 hr using the FieldSense function

Fork Opening

.........................................................17.8 mm

Temperature

Operating .............................................................-10 °C to +50 °C (+14 °F to +122 °F)
Storage ................................................................-30 °C to +60 °C (-22 °F to +140 °F)

Operating Relative Humidity

0 % to 95 %..........................................................5 °C to 30 °C (41 °F to 86 °F)
0 % to 75 %..........................................................30 °C to 40 °C (86 °F to 104 °F)
0 % to 45 %..........................................................40 °C to 50 °C (104 °F to 122 °F)

Altitude

Operating .............................................................2000 m
Storage ................................................................10 000 m

Temperature Coefficient

.......................................0.1 x (specified accuracy) / °C for <18°C or >28 °C (<64.4 °F or >82.4 °F)

Safety

General ................................................................IEC 61010-1: Pollution Degree 2
Measurement .......................................................IEC 61010-2-032, IEC 61010-2-033

T6-600 ..............................................................CAT III 600 V
T6-1000 ............................................................CAT IV 600 V / CAT III 1000 V 

Electromagnetic Compatibility (EMC)

International .........................................................IEC 61326-1: Portable Electromagnetic Environment

CISPR 11: Group 1, Class A

Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio 

frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those 
directly connected to a low-voltage power supply network that supplies buildings used for 
domestic purposes. There may be potential difficulties in ensuring electromagnetic 
compatibility in other environments due to conducted and radiated disturbances. 
Caution: This equipment is not intended for use in residential environments and may not 
provide adequate protection to radio reception in such environments.

Korea (KCC)........................................................Class A Equipment (Industrial Broadcasting & Communication Equipment)

Class A: Equipment meets requirements for industrial electromagnetic wave equipment 
and the seller or user should take notice of it. This equipment is intended for use in 
business environments and not to be used in homes.

USA (FCC) ..........................................................47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.

Model T6-1000

Model T6-600

Function

Requires Test 

Leads

Range

Range

Resolution

Accuracy

[1]

FieldSense 

 

ac voltage true-rms

No

1000 V

600 V

1 V

±(3 % + 3 counts)
45 Hz to 66 Hz

[2][3]

FieldSense 

 

ac current true-rms

No

200.0 A

200.0 A

0.1 A

±(3 % + 3 counts)
45 Hz to 66 Hz

FieldSense 

Frequency (Hz)

No

45 Hz to 66 Hz

1 Hz

±(1 % + 2 counts)

[3]

Volts ac true-rms

Yes

1000 V

600 V

1 V

±(1.5 % + 2 counts)
45 Hz to 66 Hz

Volts dc

Yes

1000 V

600 V

1 V

±(1 % + 2 counts)

Resistance

Yes

2000 

Ω

2000 

Ω

Ω

±(1 % + 2 counts)

Yes

20.00 k

Ω

0.01 k

Ω

Yes

100.0 k

Ω

0.1 k

Ω

[1] Accuracy: ±( [ % of reading] + [number of least significant digits] ). Accuracy is specified for 1 year after calibration, at 18 °C to 28 °C 

 

(64 °F to 82 °F) with relative humidity to 90 %. AC measurements are ac coupled, RMS responding.

[2] Add 3 % typical without an external ground connection. External ground connection required for user wearing insulated gloves, 

standing on an insulated ladder, or otherwise insulated from earth ground.

[3] FieldSense is specified from 16 V to 100 % of range.

Содержание T6-1000

Страница 1: ...8 2018 Fluke Corporation All rights reserved All product names are trademarks of their respective companies Specifications are subject to change without notice T6 1000 T6 600 Electrical Tester Service...

Страница 2: ...ITY OR FITNESS FOR A PARTICULAR PURPOSE FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA ARISING FROM ANY CAUSE OR THEORY Since s...

Страница 3: ...ts 5 Annunciator and Hazardous Voltage Indicator 5 Voltage Function 6 FieldSense Current Function 7 FieldSense Voltage Function 9 Resistance and Continuity Functions 9 Low Battery Indicator 10 Battery...

Страница 4: ...ii T6 1000 T6 600 Service Information...

Страница 5: ...world 1 425 446 5500 Or visit Fluke s website at www fluke com To register your product visit http register fluke com To view print or download the latest manual supplement visit http us fluke com us...

Страница 6: ...rated voltage between the terminals or between each terminal and earth ground Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product probe or accessory...

Страница 7: ...s applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Measurement Category IV is applicable to test and measuring circuits con...

Страница 8: ...s Caution This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments Korea KCC Class A Equipment Industrial Broa...

Страница 9: ...in Table 3 Table 3 Annunciator Tests 4 Set the calibrator to the voltage and frequency listed in Table 4 Step 1 5 Apply the voltage to the Product and verify the readings are within the display limits...

Страница 10: ...readings for the remaining steps in Table 6 T6 600 T6 1000 Step Calibrator Output Display Limits AC Annunciator ON Calibrator Output Display Limits AC Annunciator ON 1 200 V 60 Hz 195 V to 205 V 200...

Страница 11: ...the SHIFT yellow button on the Product to switch to AC Current mode and Frequency mode T6 1000 only 2 Connect the 10 50 turn coil to the calibrator AUX HI and AUX LO jacks 3 Connect the BLACK lead of...

Страница 12: ...hat the Tester reading is within the display limits 11 Apply the test values shown in the remaining steps of Table 8 and verify the Tester readings are within the display limits Table 8 FieldSense Hig...

Страница 13: ...the tester to switch to the AC voltage mode 8 Place the Product current fork into position on the PRV240FS 9 Place the BLACK test lead of the tester into the Test Lead Ground Contact of the PRV240FS...

Страница 14: ...sconnect the calibrator from the Tester 13 Reinstall the Tester batteries Battery Door Circuit To verify the battery door circuit operates correctly Note During this test do not use a sharp test probe...

Страница 15: ...ts Mild soap may be used Battery Replacement W Caution For safe operation and maintenance of the Product Remove batteries to prevent battery leakage and damage to the Product if it is not used for an...

Страница 16: ...able 10 To order parts and accessories see How to Contact Fluke Table 10 Replacement Parts Ref Description Fluke Part or Model Number Battery Door 4944370 Battery Door Screw 1618579 Test Lead Assembly...

Отзывы: