background image

T6-1000/T6-600

Service Information

12

Replacement Parts

Replacement parts and accessories are listed in Table 10

.

 To order parts and accessories, see 

How to Contact Fluke

.

Table 10. Replacement Parts

Ref.

Description

Fluke Part 

or Model Number

Battery Door

4944370

Battery Door Screw

1618579

Test Lead Assembly (T5-RLS). Replace only with Fluke double-insulated 
leads (

T

)

4462973

TP1 Probe Set, Flat-Blade

648128

1

2

3

4

Содержание T6-1000

Страница 1: ...8 2018 Fluke Corporation All rights reserved All product names are trademarks of their respective companies Specifications are subject to change without notice T6 1000 T6 600 Electrical Tester Service...

Страница 2: ...ITY OR FITNESS FOR A PARTICULAR PURPOSE FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA ARISING FROM ANY CAUSE OR THEORY Since s...

Страница 3: ...ts 5 Annunciator and Hazardous Voltage Indicator 5 Voltage Function 6 FieldSense Current Function 7 FieldSense Voltage Function 9 Resistance and Continuity Functions 9 Low Battery Indicator 10 Battery...

Страница 4: ...ii T6 1000 T6 600 Service Information...

Страница 5: ...world 1 425 446 5500 Or visit Fluke s website at www fluke com To register your product visit http register fluke com To view print or download the latest manual supplement visit http us fluke com us...

Страница 6: ...rated voltage between the terminals or between each terminal and earth ground Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product probe or accessory...

Страница 7: ...s applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Measurement Category IV is applicable to test and measuring circuits con...

Страница 8: ...s Caution This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments Korea KCC Class A Equipment Industrial Broa...

Страница 9: ...in Table 3 Table 3 Annunciator Tests 4 Set the calibrator to the voltage and frequency listed in Table 4 Step 1 5 Apply the voltage to the Product and verify the readings are within the display limits...

Страница 10: ...readings for the remaining steps in Table 6 T6 600 T6 1000 Step Calibrator Output Display Limits AC Annunciator ON Calibrator Output Display Limits AC Annunciator ON 1 200 V 60 Hz 195 V to 205 V 200...

Страница 11: ...the SHIFT yellow button on the Product to switch to AC Current mode and Frequency mode T6 1000 only 2 Connect the 10 50 turn coil to the calibrator AUX HI and AUX LO jacks 3 Connect the BLACK lead of...

Страница 12: ...hat the Tester reading is within the display limits 11 Apply the test values shown in the remaining steps of Table 8 and verify the Tester readings are within the display limits Table 8 FieldSense Hig...

Страница 13: ...the tester to switch to the AC voltage mode 8 Place the Product current fork into position on the PRV240FS 9 Place the BLACK test lead of the tester into the Test Lead Ground Contact of the PRV240FS...

Страница 14: ...sconnect the calibrator from the Tester 13 Reinstall the Tester batteries Battery Door Circuit To verify the battery door circuit operates correctly Note During this test do not use a sharp test probe...

Страница 15: ...ts Mild soap may be used Battery Replacement W Caution For safe operation and maintenance of the Product Remove batteries to prevent battery leakage and damage to the Product if it is not used for an...

Страница 16: ...able 10 To order parts and accessories see How to Contact Fluke Table 10 Replacement Parts Ref Description Fluke Part or Model Number Battery Door 4944370 Battery Door Screw 1618579 Test Lead Assembly...