Electrical Tester
Required Equipment
5
Required Equipment
Table 2 is a list of equipment requirements for the performance tests of the Product
.
Table 2. Required Equipment
Performance Tests
Fluke recommends a 1-year cycle for the Performance Tests. If the Product fails any test, return the Product to a Fluke Service
Center or Authorized Service Partner. See
How to Contact Fluke
for more information.
Before you start the Performance Tests, make sure that all test equipment has adequate warm-up time and is allowed to stabilize in
an environment of 23 °C ±5 °C.
Annunciator and Hazardous Voltage Indicator
Perform these tests as follows:
1.
Set the calibrator NORMAL output to the test voltage listed in Table 3, Step 1.
2.
Set the Product to the voltage (Auto Select) measurement mode.
3.
Apply the voltage to the Product and verify the display reads as shown in Table 3.
Table 3. Annunciator Tests
4.
Set the calibrator to the voltage and frequency listed in Table 4, Step 1.
5.
Apply the voltage to the Product and verify the readings are within the display limits shown in Table 4.
Table 4. Hazardous Voltage Indicator and HOLD Function Tests
Equipment
Model
Multi-Product Calibrator
5522A, 5502A/5502E, or equivalent
Precision Multimeter
8846A
DC Power Supply
0 V to 5 V adjustable
Proving Unit
PRV240FS
Test Coil
9100-200 10/50-Turn Coil (Fluke PN 1583963)
Step
Calibrator Output
Tester Display/Indicator
1
5 V dc
DC annunciator ON
2
-5 V dc
DC annunciator ON
3
5 V ac, 60 Hz
AC annunciator ON
Step
Calibrator Output
Tester Display/Indicator
1
12 V, 60 Hz
Reading: 12 V ac ±2 V
Hazardous voltage:
X
LED is OFF /
Y
on LCD is OFF
2
38 V dc
Reading: 38 V dc ±2 V
Hazardous voltage:
X
LED is ON /
Y
on LCD is ON
3
12 V, 60 Hz
Press
HOLD
on the Tester
Reading: 12 V ac ±2 V
Hazardous voltage:
X
LED is OFF /
Y
on LCD is OFF
HOLD
indicator ON
4
38 V dc
Reading: 38 V dc ±2 V
Hazardous voltage:
X
LED is ON /
Y
on LCD is ON
HOLD
indicator ON