Electrical Tester

Required Equipment

5

Required Equipment

Table 2 is a list of equipment requirements for the performance tests of the Product

.

Table 2. Required Equipment

Performance Tests

Fluke recommends a 1-year cycle for the Performance Tests. If the Product fails any test, return the Product to a Fluke Service 
Center or Authorized Service Partner. See 

How to Contact Fluke

 for more information.

Before you start the Performance Tests, make sure that all test equipment has adequate warm-up time and is allowed to stabilize in 
an environment of 23 °C ±5 °C.

Annunciator and Hazardous Voltage Indicator

Perform these tests as follows:

1.

Set the calibrator NORMAL output to the test voltage listed in Table 3, Step 1.

2.

Set the Product to the voltage (Auto Select) measurement mode.

3.

Apply the voltage to the Product and verify the display reads as shown in Table 3.

Table 3. Annunciator Tests

4.

Set the calibrator to the voltage and frequency listed in Table 4, Step 1.

5.

Apply the voltage to the Product and verify the readings are within the display limits shown in Table 4.

Table 4. Hazardous Voltage Indicator and HOLD Function Tests

Equipment

Model

Multi-Product Calibrator

5522A, 5502A/5502E, or equivalent

Precision Multimeter

8846A

DC Power Supply

0 V to 5 V adjustable

Proving Unit

PRV240FS

Test Coil

9100-200 10/50-Turn Coil (Fluke PN 1583963)

Step

Calibrator Output

Tester Display/Indicator

1

5 V dc

DC annunciator ON

2

-5 V dc

DC annunciator ON

3

5 V ac, 60 Hz

AC annunciator ON

Step

Calibrator Output

Tester Display/Indicator

1

12 V, 60 Hz

Reading: 12 V ac ±2 V
Hazardous voltage: 

X

 LED is OFF / 

Y

 on LCD is OFF

2

38 V dc

Reading: 38 V dc ±2 V
Hazardous voltage: 

X

 LED is ON / 

Y

 on LCD is ON

3

12 V, 60 Hz

Press 

HOLD

 on the Tester

Reading: 12 V ac ±2 V
Hazardous voltage: 

X

 LED is OFF / 

Y

 on LCD is OFF

HOLD

 indicator ON

4

38 V dc

Reading: 38 V dc ±2 V
Hazardous voltage: 

X

 LED is ON / 

Y

 on LCD is ON

HOLD

 indicator ON

Содержание T6-1000

Страница 1: ...8 2018 Fluke Corporation All rights reserved All product names are trademarks of their respective companies Specifications are subject to change without notice T6 1000 T6 600 Electrical Tester Service...

Страница 2: ...ITY OR FITNESS FOR A PARTICULAR PURPOSE FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA ARISING FROM ANY CAUSE OR THEORY Since s...

Страница 3: ...ts 5 Annunciator and Hazardous Voltage Indicator 5 Voltage Function 6 FieldSense Current Function 7 FieldSense Voltage Function 9 Resistance and Continuity Functions 9 Low Battery Indicator 10 Battery...

Страница 4: ...ii T6 1000 T6 600 Service Information...

Страница 5: ...world 1 425 446 5500 Or visit Fluke s website at www fluke com To register your product visit http register fluke com To view print or download the latest manual supplement visit http us fluke com us...

Страница 6: ...rated voltage between the terminals or between each terminal and earth ground Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product probe or accessory...

Страница 7: ...s applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Measurement Category IV is applicable to test and measuring circuits con...

Страница 8: ...s Caution This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments Korea KCC Class A Equipment Industrial Broa...

Страница 9: ...in Table 3 Table 3 Annunciator Tests 4 Set the calibrator to the voltage and frequency listed in Table 4 Step 1 5 Apply the voltage to the Product and verify the readings are within the display limits...

Страница 10: ...readings for the remaining steps in Table 6 T6 600 T6 1000 Step Calibrator Output Display Limits AC Annunciator ON Calibrator Output Display Limits AC Annunciator ON 1 200 V 60 Hz 195 V to 205 V 200...

Страница 11: ...the SHIFT yellow button on the Product to switch to AC Current mode and Frequency mode T6 1000 only 2 Connect the 10 50 turn coil to the calibrator AUX HI and AUX LO jacks 3 Connect the BLACK lead of...

Страница 12: ...hat the Tester reading is within the display limits 11 Apply the test values shown in the remaining steps of Table 8 and verify the Tester readings are within the display limits Table 8 FieldSense Hig...

Страница 13: ...the tester to switch to the AC voltage mode 8 Place the Product current fork into position on the PRV240FS 9 Place the BLACK test lead of the tester into the Test Lead Ground Contact of the PRV240FS...

Страница 14: ...sconnect the calibrator from the Tester 13 Reinstall the Tester batteries Battery Door Circuit To verify the battery door circuit operates correctly Note During this test do not use a sharp test probe...

Страница 15: ...ts Mild soap may be used Battery Replacement W Caution For safe operation and maintenance of the Product Remove batteries to prevent battery leakage and damage to the Product if it is not used for an...

Страница 16: ...able 10 To order parts and accessories see How to Contact Fluke Table 10 Replacement Parts Ref Description Fluke Part or Model Number Battery Door 4944370 Battery Door Screw 1618579 Test Lead Assembly...

Отзывы: