3 - 3
OMNIS
CANNER
User Guide
Scanlink Tools
support the following:
• Upload test results
• Storage of test results to files (Scanlink proprietary format)
• Print test reports
• Export results in CSV format
• Setup Projects
• Edit Autotest Information
Setting the Reference for Fiber Optic Cabling
Note
To ensure consistent accurate measurements,
the reference value must be set each time the
connector adapter or the test configuration is
changed. Since the reference value expires
each day at midnight, a reference must be
established least once each calendar day that
the OMNIF
IBER
is used.
1. Attach the connector adapters to each of the
OMNIF
IBER
RX ports.
2. To set the reference value, connect the
OMNIF
IBER
MAIN
and OMNIF
IBER
REMOTE
using
launch cables (jumpers) and couplers. See
Chapter 8 - Fiber Optic Reference Methods for
further information.
3. Press
(Setup) and then select Set
Reference.
4. The
CONNECT JUMPERS
screen will display
how the jumpers need to be connected so that
the reference value can be recorded.
5. Connect the jumpers as shown and press
(Continue) to collect the reference data
while the
SET REFERENCE
screen is displayed.
OMNIS
CANNER
will store a separate reference value for each wavelength.
Содержание OMNIFiber
Страница 1: ...OMNIScanner 2 OMNIFiber ...
Страница 9: ...OMNISCANNER User Guide TOC 6 ...
Страница 25: ...OMNISCANNER User Guide 1 16 ...
Страница 33: ...OMNISCANNER User Guide 2 8 ...
Страница 41: ...OMNISCANNER User Guide 2 16 ...
Страница 103: ...OMNISCANNER User Guide 7 6 ...
Страница 111: ...OMNISCANNER User Guide 9 2 ...
Страница 129: ...OMNISCANNER User Guide A 8 ...
Страница 133: ...OMNISCANNER User Guide C 2 ...
Страница 136: ......