W values are only meaningful with ITS-90 and IPTS-68 conversions. For
ITS-90, W refers to the ratio of measured resistance to the resistance at the tri-
ple point of water. For IPTS-68, W refers to the ratio of measured resistance to
the resistance at 0°C.
Measurement with a particular probe can be forced to be displayed in
Ω
, W, or
ratio using CONVERSION in EDIT PROBE (see Section 7.4.3.2).
7.6.2
Parameter Menu
The PARAM MENU contains functions for saving or loading all of the operat-
ing parameters to or from the disk and for setting the date and time. The param-
eter sub-menu includes the functions TIME, SAVE PARAMS, LOAD
PARAMS, RESET PARAMS and SCREEN PARAMS.
7.6.2.1
Time
The current time is always shown at the bottom of the data type screen. The
time can also be recorded with measurements that are transmitted over the com-
munications channels or saved to disk. Access to the TIME function can be
locked out as explained in Section 7.6.5.4.
The TIME soft-key lets you set the time and date. A window appears showing
the hour, minute, month and so on. You can move to any item and change it.
You can set the clock to one of two modes: 12-hour mode (12:00 a.m. to 11:59
p.m.) and 24-hour mode (0:00 to 23:59). If you will be transmitting
time-stamped measurements to a computer it may be more convenient to use
the 24-hour format. Press Enter after you make any changes to a parameter.
NOTE:
You may need to set the date again after January 1, 2000.
7.6.2.2
Save Parameters
With the SAVE PARAMS soft-key you can save all of the operating parameters
to disk. It is a good idea to keep a backup of the parameters in case they are lost
or inadvertently changed. You can also keep different sets of parameters on
disk for different applications. All of the instrument’s parameters are saved to
one file on the disk. These include all sampling variables, probe data, display
configurations, communications parameters, and reference resistor calibration
values. The file is in a special binary format for efficient storage. As a result it
cannot be easily edited on a computer.
When you use the SAVE PARAMS function, a window will appear requesting
you to enter a file number between 0 and 9. Data will be saved to a file named
“SETUPn.CFG” where n is the number you specify. Press Enter twice to save
the parameters. If a file with this name already exists on the disk, you will be
asked if you want to overwrite it.
1575/1590 Thermometer Readout
User’s Guide
84
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......