3
Specifications and Environmental
Conditions
3.1
Specifications
Specification
1575
1590
Resistance Measurement Range
0 to 500 k
Ω
Resistance Ratio Accuracy
0 to .25
Ω
input (1
Ω
refr, 10 mA)
0.00001
Ω
0.000005
Ω
0.25 to 4
Ω
input (1
Ω
refr, 10 mA)
40 ppm
20 ppm
2.5 to 40
Ω
input (10
Ω
refr, 3 mA)
20 ppm
5 ppm
0 to 25
Ω
input (100
Ω
refr, 1 mA)
0.0001
Ω
0.000025
Ω
25 to 400
Ω
input (100
Ω
refr, 1 mA)
4 ppm
1 ppm
400 to 1000
Ω
input (100
Ω
refr, 0.1 mA)
20 ppm
4 ppm
0 to 2.5 k
Ω
input (10 k
Ω
refr, 0.01 mA)
0.025
Ω
0.012
Ω
2.5 to 40 k
Ω
input (10 k
Ω
refr, 0.01 mA)
10 ppm
5 ppm
40 to 100 k
Ω
input (10 k
Ω
refr, 0.005 mA)
50 ppm
25 ppm
100 to 500 k
Ω
input (10 k
Ω
refr, 0.002 mA)
200 ppm
120 ppm
Resistance Accuracy
0 to .25
Ω
input (1
Ω
refr, 10 mA)
0.000025
Ω
0.00001
Ω
0.25 to 4
Ω
input (1
Ω
refr, 10 mA)
100 ppm
40 ppm
2.5 to 40
Ω
input 10
Ω
refr, 3 mA)
40 ppm
20 ppm
0 to 25
Ω
input (100
Ω
refr, 1 mA)
0.0002
Ω
0.00015
Ω
25 to 400
Ω
input (100
Ω
refr, 1 mA)
8 ppm
6 ppm
400 to 1000
Ω
input (100
Ω
refr, 0.1 mA)
25 ppm
10 ppm
0 to 2.5 k
Ω
input (10 k
Ω
refr, 0.01 mA)
0.05
Ω
0.025
Ω
2.5 to 40 k
Ω
input (10 k
Ω
refr, 0.01 mA)
20 ppm
10 ppm
40 to 100 k
Ω
input (10 k
Ω
refr, 0.005 mA)
60 ppm
30 ppm
100 to 500 k
Ω
input (10 k
Ω
refr, 0.002 mA)
200 ppm
120 ppm
Typical Temperature Accuracy, External Reference
0.25
Ω
SPRT (0
°
C, 1
Ω
refr, 10 mA)
0.01°C
0.005°C
2.5
Ω
SPRT (0
°
C, 10
Ω
refr, 3 mA)
0.005°C
0.0013°C
25
Ω
SPRT (0
°
C, 100
Ω
refr, 1 mA)
0.001°C
0.00025°C
100
Ω
SPRT (0
°
C, 100
Ω
refr, 1 mA)
0.001
°
C
0.00025
°
C
10 k
Ω
thermistor (25
°
C, 10 k
Ω
refr, 0.01 mA)
0.00025°C
0.00013°C
Typical Temperature Accuracy, Internal Reference
0.25
Ω
SPRT (0
°
C, 1
Ω
refr, 10 mA)
0.025°C
0.01°C
2.5
Ω
SPRT (0
°
C, 10
Ω
refr, 3 mA)
0.01°C
0.005°C
25
Ω
SPRT (0
°
C, 100
Ω
refr, 1 mA)
0.002°C
0.0015°C
13
3 Specifications and Environmental Conditions
Specifications
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......