8.2
IEEE-488 Communications
The IEEE-488 (GPIB) interface works similarly to the serial interface. One ad-
vantage of the IEEE-488 over serial is that many devices can be connected to a
single bus. Also, it is faster than serial communications.
8.2.1
Setup
To set up the IEEE-488 communications, first connect an IEEE-488 standard
cable to the back of the 1575/1590. Next, set the device address and, if neces-
sary, the termination character. These parameters are programmed with the
IEEE488 soft-key in the SYSTEM - COMM MENU (see Section 7.6.4.2). You
can also set the time or date stamp option for transmitted measurements. Com-
mands can be sent from the bus controller to read the temperature or program
parameters. The same commands are used with IEEE-488 that are used with se-
rial communications. These are listed in Section 8.3.
8.2.2
IEEE-488 Interface Functions
Specific IEEE-488 interface features and commands that are implemented with
the 1575/1590 are described below. The codes that describe the GPIB function-
ality of the 1575/1590 are: SH1, AH1, T6, L4, SR1, RL1, PP1, DC1, DT1, C0,
and E2.
8.2.2.1
Local Lockout
You can use the local lockout (LLO) GPIB command to disable the front panel
buttons while remotely controlling the 1575/1590. When the 1575/1590 re-
ceives the LLO command, the soft-key menu will disappear and the message
“LOCAL LOCKOUT” will appear in the upper right corner of the display. Lo-
cal mode can be restored by sending the go-to-local (GTL) command. The
1575/1590 will always power up in local mode.
8.2.2.2
Trigger
You can use the trigger (GET) GPIB command to start a new measurement cy-
cle. If the 1575/1590 is in the process of measuring when it receives the GET
101
8 Communications Interface
IEEE-488 Communications
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......