7.6.5.2
Set Internal
The SET INT soft-key is used to set the calibration values of the internal refer-
ence resistors when the 1575/1590 is calibrated. This function also allows you
to enable or disable the temperature stabilization oven for the reference resis-
tors. These settings are critical to the accuracy of the instrument so they should
not be changed except when necessary to calibrate the 1575/1590. (For an ex-
planation of the calibration procedure, see Section 9.) Access to the SET INT
function may be locked out as explained in Section 7.6.5.4.
Normally, the oven should be left on since it helps to maintain the accuracy of
the internal reference resistors. The thermometer is calibrated with the oven on.
The oven may be turned off if the absolute accuracy of the reference resistors is
not critical and the warm-up time after power on must be minimized.
CAUTION
: Do not arbitrarily change the calibration values of the refer-
ence resistors as this may affect the accuracy of the instrument.
7.6.5.3
Calibrate Reference
The CAL REFR soft-key is used to calibrate the values of the internal reference
resistors. Correct calibration of the reference resistors is critical to the accuracy
of the instrument so this should not be done except by a qualified technician.
(For an explanation of the calibration procedure see Section 9.) Access to the
CAL REFR function may be locked out as explained in Section 7.6.5.4.
7.6.5.4
Security Lock-out
The SECURITY function in the SYSTEM CAL menu will allow you lock-out
access to certain parameters, menus, and functions. This provides protection so
that critical settings that may affect the accuracy and proper operation of the
1575/1590 cannot be inadvertently changed. The following groups of settings
can be independently locked out.
1.
Reference resistor calibration values (SET INT) and calibration proce-
dure (CAL REFR).
2.
Probe characterization and setup parameters and probe calibration func-
tions (PROBE MENU).
3.
Filter and sampling parameters (DIGITAL FILTER and SAMPLE
TIMING).
4.
System units and time and date (UNITS and TIME).
Any menu or function that is locked out cannot be accessed without providing
the correct four-digit password. The user can specify the password. The default
is “1575" or ”1590". The security option window itself cannot be accessed
without entering the password. When access to any locked-out screen is at-
tempted, a message will appear and prompt the user to enter the password. If
the password is entered correctly access will be allowed.
95
7 Front Panel Operation
System Menu
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......