IEEE488 soft-key. These include the address, termination character, and time
and date stamp. Pressing this soft-key brings up a window as shown below.
DEVICE ADDRESS sets the IEEE-488 address of the 1575/1590 Super-Ther-
mometer. To be used as a device on an IEEE-488 bus the thermometer must be
given a unique address. It can be any number from 0 to 30. It must be different
from any address used by any other devices also connected to the bus.
TERM determines the character or characters that are sent after each line of
data. You can select linefeed, carriage return, or both as the termination.
The TIME and DATE parameters determine whether the current time and date
are transmitted along with measurements (in response to the “tem” command,
see Section 8.3). The time can be set in 12-hour or 24-hour time format (Sec-
tion 7.6.2.1).
7.6.4.3
Printer
With the printer (parallel) interface, you can connect a printer to the thermome-
ter and have it record measurements on paper. Data sampled to the printer port
is formatted the same as data recorded to disk. Figure 20 shows an example of
how the data might look. The example is with the PERIOD set to AUTO and
TIME and DATE ON (see below). The channel mode is set to ALT N with N=1
(see Section 7.2.3). Each measurement value is preceded by the channel
number.
To transmit data to a printer, you must set the period. You can also set the line-
1575/1590 Thermometer Readout
User’s Guide
90
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......