1575/1590 Thermometer Readout
User’s Guide
106
Command
Description
Example
po[ffset](p)[=n]
read or set three-wire lead offset for probe
po(16)=0.2857
rl[90](p)[=n]
read or set ITS-90 low range selection
rl90(16)=4
rh[90](p)[=n]
read or set ITS-90 high range selection
rh90(16)=7
r9[0](p)[=n]
read or set ITS-90 R(TPW) probe constant
r90(16)=25.48635
a9[0](p,r)[=n]
read or set ITS-90 a probe constants for range
a90(16,7)=-3.4837E-4
b9[0](p,r)[=n]
read or set ITS-90 b probe constants for range
b90(16,7)=1.5385E-5
c9[0](p,r[,i])[=n]
read or set ITS-90 c or ci probe constants for range
c90(16,7)=1.3820E-6
d9[0](p)[=n]
read or set ITS-90 d constant
d90(16)=4.876534E-9
r6[8](p)[=n]
read or set IPTS-68 R(0) constant
r68(16)=100.037465
al6[8](p)[=n]
read or set IPTS-68 alpha constant
al68(16)=0.00392583
de6[8](p)[=n]
read or set IPTS-68 delta constant
de68(16)=1.495873
a6[8](p,r)[=n]
read or set IPTS-68 a constant for range
a68(16,4)=-3.8765E-6
b6[8](p,r)[=n]
read or set IPTS-68 b constant for range
b68(16,3)=8.2094E-8
c6[8](p,r)[=n]
read or set IPTS-68 c constant for range
c68(16,4)=1.4930E-7
d6[8](p,r)[=n]
read or set IPTS-68 d constant for range
d68(16,2)=-3.3654E-6
scal[e](p)[=9[0]/6[8]]
read or select scale conversion, ITS-90 or IPTS-68
scale(16)=68
rc[vd](p)[=n]
read or set Callendar Van-Dusen R(0) constant
rcvd(16)=100.0
alc[vd](p)[=n]
read or set Callendar Van-Dusen alpha constant
alcvd(16)=0.00385
dec[vd](p)[=n]
read or set Callendar Van-Dusen delta constant
decvd(16)=1.507
bec[vd](p)[=n]
read or set Callendar Van-Dusen beta constant
becvd(16)=0.111
ap[oly](p,i)[=n]
read or set polynomial coefficients
ap(16,0)=31.9438576
at[herm](p,i)[=n]
read or set thermistor coefficients ai [T(R)]
at(16,0)=0.00368732
bt[herm](p,i)[=n]
read or set thermistor coefficients bi [R(T)]
bt(16,0)=-4.4728857
Display Commands
drt[emperature][=n]
read or set digits of resolution for temperature (C, K, F)
drt=4
dro[hms][=n]
read or set digits of resolution for resistance (W)
dro=6
drr[atio][=n]
read or set digits of resolution for ratio (r, W)
drr=8
dn[umber][=n]
read or set display number
dn=16
dt[ype][=d[ata]/g[raph]]
read or set display type
dt=g
dfp[arameter](d,f)[=n]
read or set display field parameter (see Section 5.6.3)
dfp(16,5)=3
dfi[ndex](d,i)[=n]
read or set display field memory/history index
dfi(16,5)=100
grhm[ode](d)[=a[uto]/ s[et]]
read or set mode of high limit for graph display
grhm(16)=s
grlm[ode](d)[=a[uto]/s[et]]
read or set mode of low limit for graph display
grlm(16)=s
grhv[alue](d)[=n]
read or set value of high limit for graph display
grhv(16)=100
grlv[alue](d)[=n]
read or set value of low limit for graph display
grlv(16)=0
Table 12
Interface Commands continued
Содержание Hart Scientific 1575 Super-Thermometer
Страница 1: ...1575 1590 Thermometer Readout User s Guide Rev 5B2802 Hart Scientific ...
Страница 10: ......
Страница 22: ......
Страница 26: ......
Страница 30: ......
Страница 38: ......
Страница 46: ......
Страница 108: ......
Страница 122: ......
Страница 134: ......
Страница 146: ......
Страница 152: ......