![Fluke 9100A Series Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/fluke/9100a-series/9100a-series_manual_2309313054.webp)
9100A-017
6-10
If the groups line or the display line are modified after data fields
have been entered into the file, the data fields change to match
the modifications. For example, if the following vector file was
loaded:
GROUPS [40-9] [8-1]
DISPLAY HEX, BIN
!TIME DATA
1 $401007FD 1111XXXX
2 $FFFFFFFF 0000XXXX
3 $00000000 1100XX11
and the groups line was changed to read:
GROUPS [40-13] [12-1]
the vector file would change to:
GROUPS [40-13] [12-1]
DISPLAY HEX, BIN
!TIME DATA
1 $401007F 11011111XXXX
2 $FFFFFFF 11110000XXXX
3 $0000000 00001100XX11
If the display line was then changed to read:
DISPLAY BIN, HEX
the vector file would change to:
GROUPS [40-13] [12-1]
DISPLAY BIN, HEX
!TIME DATA
1 0100000000010000000001111111 $DFX
2 1111111111111111111111111111 $F0X
3 0000000000000000000000000000 *00001100XX11
Содержание 9100A Series
Страница 6: ...vi ...
Страница 8: ...viii ...
Страница 10: ...x ...
Страница 14: ...9100A 017 1 4 ...
Страница 24: ...9100A 017 3 6 ...
Страница 44: ...9100A 017 5 4 ...
Страница 58: ...9100A 017 6 14 ...
Страница 83: ...A 1 Appendix A New TL 1 Commands ...
Страница 84: ...9100A 017 A 2 ...
Страница 87: ...clockfreq 3 For More Information The Overview Of TL 1 section of the Programmer s Manual ...
Страница 88: ...clockfreq 4 ...
Страница 91: ...drivepoll 3 For More Information The Overview Of TL 1 section of the Programmer s Manual ...
Страница 92: ...drivepoll 4 ...
Страница 104: ...vectordrive 4 ...
Страница 107: ...vectorload 3 For More Information The Overview Of TL 1 section of the Programmer s Manual ...
Страница 108: ...vectorload 4 ...
Страница 116: ...9100A 017 C 2 ...
Страница 117: ...9100A 017 C 3 ...
Страница 118: ...9100A 017 C 4 ...