Duplex Mode
The duplex mode can be set to full duplex or half duplex. With full duplex
,
any commands received by
the calibrator with the serial interface will be immediately echoed or transmitted back to the device of
origin. With half duplex
,
the commands will be executed but not echoed. The duplex mode parameter is
indicated by,
dUPL
Flashes dUPL and then displays the setting
d=FULL
Current duplex mode setting
The mode may be changed with
UP
or
DOWN
and
SET
.
d=HALF
New duplex mode setting
Linefeed
Linefeed enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after
transmission of any carriage-return.
LF
Flashes LF and then displays the setting
LF= On
Current linefeed setting
The mode may be changed with
UP
or
DOWN
and
SET
.
LF= OFF
New linefeed setting
Calibration Parameters
The controller gives access to a number of the Product calibration constants: R0, ALPHA, DELTA, and
BETA (cold side only). These values are set at the factory and should not be altered. The correct
values are important to the accuracy and proper and safe operation of the Product. Access to these
parameters is available so that if the controller memory fails
,
the user may restore these values to the
factory settings. Keep a list of these constants and their settings with the manual. These constants and
their settings are on the Report of Calibration that is shipped with the instrument.
W
Caution
Do not change the values of the Product calibration constants from the factory
-
set
values. The correct setting of these parameters is important to the safety and proper
operation of the Product.
The calibration parameters menu is indicated by:
CAL
Calibration parameters menu
Push
SET
five times to enter the menu.