8508A
Service Manual
3-40
Table 3-10. Resistance Verification Points and Sequence (cont)
Tolerance
1
8508A Range
Resistance
HiV
Ω
2
0 G
Ω
Perform input zero operation to calibrator range locked zero output
20 M
Ω
10 M
Ω
±
0.0000202 M
Ω
0 G
Ω
Perform input zero operation to calibrator range locked zero output
200 M
Ω
100 M
Ω
±
0.0000870 M
Ω
0 G
Ω
Perform input zero operation to calibrator range locked zero output
2 G
Ω
1 G
Ω
±
0.0003150 G
Ω
0 G
Ω
Perform input zero operation to calibrator range locked zero output
20 G
Ω
1 G
Ω
±
0.012675 G
Ω
[1]
Based on 365 day Tcal
±
1
°
C 99 %CL Absolute specifications. See comments in Applicable Verification Tolerance section above
regarding the impact of calibration uncertainties and use of other specifications.
[2]
Ranges require separate input zero operation in each resistance mode.
DC Current Checks
Equipment Configuration
1.
Ensure that the calibrator output is set to
STANDBY
and configured for Internal
Guard (
EXT GRD
is deselected).
2.
Connect the equipment as described for Routine Calibration.
3.
Press the
DCI
key
4.
Press the
CONFIG
key. Select
Filt
and
RESL7
.
5.
Confirm that the Multimeter is configured for internal guard (External Guard is
deselected) by checking that the
Ext Grd
legend does not appear on the left-hand
display.
Procedure
Repeat the following sequence for each range, starting with the 200
μ
A range and
working up to the 20A range, as listed in Table 3-11. An input zero operation is
performed on each Multimeter range prior to performing the range verification
measurements with the calibrator set to zero by range locking the calibrator in the range
from which the output will be obtained.
1.
Select the required Multimeter range.
2.
On the calibrator set the output to zero. Select OPERATE.
3.
On the Multimeter press the
INPUT
key, and select
Zero Rng
to initiate the zeroing
operation.
The
Busy
legend will appear, and then extinguish when the zero operation has
completed.
4.
On the calibrator select the value required for verification in the positive polarity.
5.
Note the Multimeter’s reading. Compare the result with the allowable tolerance,
taking into account any known error from nominal of the calibrator for this output.
Содержание 8508A
Страница 6: ...8508A Service Manual iv ...
Страница 8: ...8508A Service Manual vi ...
Страница 10: ...8508A Service Manual viii ...
Страница 12: ...8508A Service Manual 1 2 ...
Страница 32: ...8508A Service Manual 1 22 ...
Страница 34: ...8508A Service Manual 2 2 ...
Страница 40: ...8508A Service Manual 2 8 ...
Страница 84: ...8508A Service Manual 3 44 ...
Страница 88: ...8508A Service Manual 4 2 ...
Страница 108: ...8508A Service Manual 5 2 ...
Страница 120: ...8508A Service Manual 5 14 ...