4-13
Rear-Input Current Fuse F2 (Rear Inputs Option)
The rear input option includes a fuse (F2) to limit the current at the rear-input A terminal
to 2A. The fuse is located at the rear of the Multimeter just to the left of the current input
terminal (A). A half-twist of the fuse cap releases the fuse for inspection or replacement.
Use the fuse specified in Table 4-2 when replacement is necessary.
Table 4-2. Current Function Rear Input Fuse 2
Fuse Action
Fuse Rating
IEC 127 (UL / CSA)
Fluke Part No.
Manufacturer
& Type No.
FH
Fast acting HBC
1.6 A (2 A)
920071
Beswick S501
Self Test
This section of the manual contains a collection of the error codes which might be
generated either on the display of the Multimeter or by way of the IEEE-488 system bus.
Error codes described in this section are shown in tabular form with both the error codes
and the descriptions of the most likely faults, causes, or conditions. Typically, the error
codes contain two parts, a four digit address code and a two digit id code. However, some
error codes consist only of the four digit address. The combination for any given code is
identified by the presence or absence of the id column in the error code table.
The error codes are classified as Self-Test Errors, Fatal Errors, Execution Errors, and
Device Dependant Errors. All error conditions other than Fatal Errors are recoverable.
For help in determining what to do in the event of a self-test error, make a note of the full
error message and contact your nearest Fluke Service Center.
Self-Test Errors
When a Self-Test is run and successfully completed no error messages result and the
Multimeter is considered operational. If a self-test error does occur, an error message
beginning with
self test failed
shows on the display. A function indicator and a
test id number follow to complete the message. See Table 4-3 for descriptions of the self-
test errors that might occur.
Table 4-3. Self-Test Errors
DC VOLTS
Test No 10
Err 2101
P001
Main Ref Check
Test No 11
Err 2101
P002
6.2V Ref Check
Test No 20
Err 2201
P003
100mV Range Zero Check
Test No 21
Err 2211
P004
1V Range Zero Check
Test No 22
Err 2221
P005
10V Range Zero Check
Test No 23
Err 2231
P006
100V Range Zero Check
Test No 24
Err 2241
P007
100mV Range Positive Gain Check
Test No 25
Err 2251
P008
100mV Range Negative Gain Check
Test No 26
Err 2261
P009
1V Range Positive Gain Check
Test No 27
Err 2271
P010
1V Range Negative Gain Check
Содержание 8508A
Страница 6: ...8508A Service Manual iv ...
Страница 8: ...8508A Service Manual vi ...
Страница 10: ...8508A Service Manual viii ...
Страница 12: ...8508A Service Manual 1 2 ...
Страница 32: ...8508A Service Manual 1 22 ...
Страница 34: ...8508A Service Manual 2 2 ...
Страница 40: ...8508A Service Manual 2 8 ...
Страница 84: ...8508A Service Manual 3 44 ...
Страница 88: ...8508A Service Manual 4 2 ...
Страница 108: ...8508A Service Manual 5 2 ...
Страница 120: ...8508A Service Manual 5 14 ...