![Fluke 8508A Скачать руководство пользователя страница 101](http://html1.mh-extra.com/html/fluke/8508a/8508a_service-manual_2310167101.webp)
Maintenance
Self Test
4
4-15
AC VOLTS
Test No 80
Err 2801
P015
100mV AC Range: Zero Input Check
Test No 81
Err 2811
P016
100mV AC Range: +179mV DC Input Checks at
PreAmp Output
Test No 82
Err 2821
P017
100mV AC Range: -179mV DC Input Checks at
PreAmp Output
Test No 83
Err 2831
2832 raw
P018
1V AC Range: Zero Input Check
Test No 84
Err 2841
P019
1V AC Range: +993mV DC Input Checks at
PreAmp Output
Test No 85
Err 2851
P020
1V AC Range: -993mV DC Input Checks at PreAmp
Output
Test No 86
Err 2861
P021
1V AC Range: +993mV DC Input Checks at RMS
Converter Output
Test No 87
Err 2871
P022
1V AC Range: -993mV DC Input Checks at RMS
Converter Output
Test No 88
Err 2881
2882 raw
P023
10V AC Range: Zero Input Check
Test No 89
Err 2891
P024
10V AC Range: +9.78V DC Input Checks at
PreAmp Output
Test No 90
Err 2901
P025
10V AC Range: -9.78V DC Input Checks at PreAmp
Output
Test No 91
Err 2911
2912 raw
P026
100V AC Range: Zero Input Check
Test No 92
Err 2921
P027
100V AC Range: +9.78V DC Input Checks at
PreAmp Output
Test No 93
Err 2931
P028
100V AC Range: -9.78V DC Input Checks at
PreAmp Output
Test No 94
Err 2941
2942 raw
P029
1000V AC Range: Zero Input Check
Test No 95
Err 2951
P030
1000V AC Range: +9.78V DC Input Checks at
PreAmp Output
Test No 96
Err 2961
P031
1000V AC Range: -9.78V DC Input Checks at
PreAmp Output
Содержание 8508A
Страница 6: ...8508A Service Manual iv ...
Страница 8: ...8508A Service Manual vi ...
Страница 10: ...8508A Service Manual viii ...
Страница 12: ...8508A Service Manual 1 2 ...
Страница 32: ...8508A Service Manual 1 22 ...
Страница 34: ...8508A Service Manual 2 2 ...
Страница 40: ...8508A Service Manual 2 8 ...
Страница 84: ...8508A Service Manual 3 44 ...
Страница 88: ...8508A Service Manual 4 2 ...
Страница 108: ...8508A Service Manual 5 2 ...
Страница 120: ...8508A Service Manual 5 14 ...