8508A
Service Manual
3-38
Procedure
Perform the following sequence for each resistance mode and each range, as listed in
Table 3-10, starting with the 2
Ω
range and working up to the 2 G
Ω
range for the
Normal
Ω
and Normal
Ω
LoI resistance modes. Then continue the sequence for the Tru
Ω
and Tru
Ω
LoI modes, starting with the 2
Ω
range and working up to the 20 k
Ω
range.
Finally complete the sequence for the HiV
Ω
mode, starting with the 20 M
Ω
range and
working up to the 20 G
Ω
range. An input zero operation is performed on each Multimeter
range prior to performing the range verification measurements with the calibrator set to
zero
Ω
. A separate zero operation is required on each range for each resistance mode.
Note that EX SNS (4-wire) is not available on the 5720A 100 M
Ω
range, and the 8508-
7000K 1 G
Ω
standard resistor is used as a 2-wire device with the calibrator 2-wire
resistance zero used for the corresponding zero point.
4W
Ω
is deselected
on the
Multimeter when checking the 200 M
Ω
and 2 G
Ω
range, range zero, and range gain
points.
XW
Warning
The Multimeter can deliver a lethal electric shock when
operating in the High Voltage Resistance mode. Voltages up to
240V may appear on the terminals.
To avoid electric shock when performing the following
calibration procedures:
•
Never touch any lead or terminal unless you are absolutely
certain that no dangerous voltage is present.
•
Make sure that signal leads are in a safe condition before you
handle them in any way.
1.
Press the
Ω
key.
2.
Press the
CONFIG
key and select
Filt
,
RESL7
, and
4w
Ω
. Ensure
LoI
is deselected.
3.
Select the required Multimeter range.
4.
On the Calibrator, select zero
Ω
and Operate. Alternatively if using a standard
resistor connect the standard resistor for a 4-wire resistance zero. (For the 200 M
Ω
range and above, de-select Ex SNS on the calibrator and de-select
4W
Ω
ο
n the
Multimeter.)
5.
On the Multimeter press the
INPUT
key, and select
Zero Rng
to initiate the zeroing
operation.
The
Busy
legend will appear, and then extinguish when the zero operation has
completed.
6.
On the Calibrator, select Range Gain point Output value. Alternatively if using a
standard resistor re-connect the standard resistor to measure its resistance.
7.
Note the Multimeter’s reading. Compare the result with the allowable tolerance,
taking into account the actual value of the calibrator resistance output for this value
or the calibrated value of the resistance standard used.
8.
Repeat steps 3 – 7 for the remaining ranges of the selected resistance mode. (For the
2 G
Ω
and 20 G
Ω
ranges the 8508-7000K 1 G
Ω
standard resistor is used in place of
the calibrator. For the 200 M
Ω
range and above, de-select Ex SNS on the calibrator
and de-select
4W
Ω
ο
n the Multimeter).
9.
Set the calibrator output to
STANDBY
.
Содержание 8508A
Страница 6: ...8508A Service Manual iv ...
Страница 8: ...8508A Service Manual vi ...
Страница 10: ...8508A Service Manual viii ...
Страница 12: ...8508A Service Manual 1 2 ...
Страница 32: ...8508A Service Manual 1 22 ...
Страница 34: ...8508A Service Manual 2 2 ...
Страница 40: ...8508A Service Manual 2 8 ...
Страница 84: ...8508A Service Manual 3 44 ...
Страница 88: ...8508A Service Manual 4 2 ...
Страница 108: ...8508A Service Manual 5 2 ...
Страница 120: ...8508A Service Manual 5 14 ...