8508A
Service Manual
3-30
6.
Press the
CONFIG
key and select
Filt
,
LoI
and
4w
Ω
. Repeat steps 3 and 4.
7.
Press the
CONFIG
key and select
2w
Ω
. Repeat steps 3 and 4.
8.
Press the
CONFIG
key, select
4w
Ω
and de-select
LoI
. Repeat steps 3 and 4.
9.
Press the
CONFIG
key and select
2w
Ω
. Repeat steps 3 and 4.
10.
Press the
Ω
PLUS
key and select
Tru
Ω
. From the
Tru
Ω
menu the 2
Ω
range.
11.
Press the
CONFIG
key and select
Filt
and
LoI
. Repeat steps 3 and 4.
12.
Press the
CONFIG
key
and de-select
LoI
. Repeat steps 3 and 4.
13.
Press the
Ω
PLUS
key and select
HiV
Ω
. From the
HiV
Ω
menu the 20M
Ω
range.
14.
Press the
CONFIG
key and select
Filt
and
2w
Ω
. Repeat steps 3 and 4.
15.
Press the
CONFIG
key
and select
4w
Ω
. Repeat steps 3 and 4.
16.
Press the
DCV
key and remove the shorting devices from the input terminals.
Cal Due Date Entry and Calibration Mode Disable
On completion of all calibration operations it is recommended that the user updates the
stored
CAL
due date with the date for next recalibration. For consistency, this date should
be the same as the calibration due date shown on any calibration stickers the user may
also attach to the Multimeter. It is also recommended that a calibration seal or integrity
sticker is applied to prevent unauthorized access to the rear panel calibration enable
switch.
Set Cal Due Date
1.
Press the
CAL
key and select Quit.
2.
From the
CAL DUE ?
menu select Enter.
3.
Use the keypad to key in the calibration due date. Up to eight characters may be
entered in any format. Any character available from the keypad may be used, but it is
recommended that one of the common date formats such as dd.mm.yy or mm.dd.yy
is used for ease of interpretation.
4.
Press Enter to store the
CAL
due date.
Disable cal mode
1.
Set the rear panel Calibration Enable switch to
DISABLE
.
2.
Apply a security seal to cover the rear panel Calibration Enable switch to prevent
unauthorized access.
ACV Spot Frequency Calibration
The ACV Spot frequency mode calibration is not normally performed as part of routine
calibration. ACV Spot frequency calibration is only performed if the user intends to make
use of spot frequency mode. In spot frequency mode the accuracy of AC Voltage
measurements can be improved by calibrating the Multimeter at specific frequencies of
interest, reducing frequency response flatness errors at these points. The following
procedure provides guidance on spot frequency calibration for users wishing to make use
of Spot frequency mode.
Содержание 8508A
Страница 6: ...8508A Service Manual iv ...
Страница 8: ...8508A Service Manual vi ...
Страница 10: ...8508A Service Manual viii ...
Страница 12: ...8508A Service Manual 1 2 ...
Страница 32: ...8508A Service Manual 1 22 ...
Страница 34: ...8508A Service Manual 2 2 ...
Страница 40: ...8508A Service Manual 2 8 ...
Страница 84: ...8508A Service Manual 3 44 ...
Страница 88: ...8508A Service Manual 4 2 ...
Страница 108: ...8508A Service Manual 5 2 ...
Страница 120: ...8508A Service Manual 5 14 ...