5730A
Calibration Manual
24
Settling Time
...................................................... 5 seconds for 5730A ranges; 6 seconds for 5725A 11 A range; +1
second for amplitude or frequency range change.
Overshoot
.......................................................... <10 %
Wideband AC Voltage (Option 5700-03) Specifications
Specifications apply to the end of the cable and 50
Ω
termination used for calibration.
Range
Resolution
Absolute /
±
5
°
C from calibration temperature
30 Hz - 500 kHz
Volts
dBm
24 Hours
90 Days
180 Days
1 Year
±
(%
μ
V)
1.1 mV
-46
10 nV
0.4 + 0.4
0.5 + 0.4
0.6 + 0.4
0.8 + 2
3.3 mV
-37
10 nV
0.4 + 1
0.45 + 1
0.5 + 1
0.7 + 3
11 mV
-26
100 nV
0.2 + 4
0.35 + 4
0.5 + 4
0.7 + 8
33 mV
-17
100 nV
0.2 + 10
0.3 + 10
0.45 + 10
0.6 + 16
110 mV
-6.2
1
μ
V
0.2 + 40
0.3 + 40
0.45 + 40
0.6 + 40
330 mV
+3.4
1
μ
V
0.2 + 100
0.25 + 100
0.35 + 100
0.5 + 100
1.1 V
+14
10
μ
V
0.2 + 400
0.25 + 400
0.35 + 400
0.5 + 400
3.5 V
+24
10
μ
V
0.15 + 500
0.2 + 500
0.3 + 500
0.4 + 500
Frequency
(Hz)
Frequency
Resolution
(Hz)
Amplitude Flatness, 1 kHz Reference
Voltage Range
Temperature
Coefficient
±
ppm/
°
C
Settling Time
To Published
Specification
(Seconds)
Harmonic
Distortion
(dB)
1.1 mV
3.3 mV
>3.3 mV
±
(% floor indicated)
10 - 30
0.01
0.3
0.3
0.3
100
7
-40
30 - 119.99
0.01
0.1
0.1
0.1
100
7
-40
120 - 1.1999 k
0.1
0.1
0.1
0.1
100
5
-40
1.2 k - 11.999 k
1
0.1
0.1
0.1
100
5
-40
12 k - 119.99 k
10
0.1
0.1
0.1
100
5
-40
120 k - 1.1999 M
100
0.2 + 3
μ
V
0.1 + 3
μ
V
0.1 + 3
μ
V
100 5
-40
1.2 M - 2 M
[1]
100 k
0.2 + 3
μ
V
0.1 + 3
μ
V
0.1 + 3
μ
V
100 0.5
-40
2 M - 11.9 M
100 k
0.4 + 3
μ
V
0.3 + 3
μ
V
0.2 + 3
μ
V
100 0.5
-40
12 M - 20 M
1 M
0.6 + 3
μ
V
0.5 + 3
μ
V
0.4 + 3
μ
V
150 0.5
-34
20 M - 30 M
1 M
1.5 + 15
μ
V
1.5 + 3
μ
V
1 + 3
μ
V
300 0.5
-34
Note:
1.
For output voltages <50 % of full range in the 33 mV, 110 mV, 330 mV, 1.1 V , and 3.5 V ranges, add 0.1 % to the amplitude
flatness specification.
Additional Operating Information:
dBm reference = 50
Ω
Range boundaries are at voltage points, dBm levels are approximate.
dBm = 10 log (
mW
1
Power
) ; 0.22361 V across 50
Ω
= 1 mW or 0 dBm
Minimum Output
................................................... 300
μ
V (-57 dBm)
Frequency Specification
......................................
±
0.01 % of output
Frequency Resolution
.......................................... 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz
Overload Protection
.............................................. A short circuit on the wideband output will not result in damage. After
settling time, normal operation is restored upon removal.