Remote Commands
Remote Command Listing
5
5-21
*WAI
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Wait-to-Continue command) Prevents further remote commands from being
executed until all previous remote commands have been executed. For example, if
you send an
OUT
command, you can cause the Calibrator to wait until the output
has settled before continuing on to the next command if you follow
OUT
with a
*WAI
command. The
*WAI
command is useful with any overlapped command,
preventing the Calibrator from processing other commands until the overlapped
command is processed.
Example:
*WAI
Process all existing commands before continuing.
ZERO_MEAS
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Zero Measure command) Zeros the pressure module and the TC mV offset. See
Chapter 3 for the TC mV zero calibration procedure.
Parameters:
<value>
in current pressure units for absolute pressure modules.
Example:
ZERO_MEAS
ZERO_MEAS?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Zero Measure query) Returns the zero offset for absolute pressure modules.
Response:
<zero offset>,<units>
Example:
ZERO_MEAS?
returns
−
1.00,PSI