Maintaining the Calibrator
Performance Tests
6
6-9
Testing Thermocouple Input
For this test the CJC (cold junction compensation) must be turned off and the leads
and mini-connector must be copper and copper. Press
I
5
to turn off the CJC.
XCJC on the display indicates that the CJU is turned off. The mini-connector will
have the designation “Cu” for copper and copper. Connect the mini-connector to
the front of the 525A TC input/output connector and the other end to the 5520A
DC voltage output. See Figure 6-5. Output the simulated thermocouple
temperatures listed in Table 6-4.