Specifications
General Specifications
Warm up time
Twice the time since last warmed up, to a maximum of 30 minutes.
Settling time
Less than 5 seconds for all functions and ranges except as noted.
Standard interface
RS-232
Optional interface
IEEE-488 (GPIB)
Temperature performance
Operating
0
°
C to 50
°
C
Calibration (tcal)
18
°
C to 28
°
C
Storage
–20
°
C to 70
°
C
Electromagnetic compatibility
CE: Conforms to EN61326
Temperature coefficient
Temperature coefficient for temperatures outside tcal
±
5
°
C is
10 % of the 90 day specification (or 1 year if applicable) per
°
C
Relative humidity
Operating
<80 % to 30
°
C, <70% to 40
°
C, <40 % to 50
°
C
Storage
<95 % noncondensing
Altitude
Operating
3,050 m (10,000 ft) maximum
Nonoperating
12,200 m (40,000 ft) maximum
Safety
EN 61010 Second, ANSI/ISA-S82.01-1994,
CAN/CSA-C22.2 No. 1010.1-92, NRTL
Analog low isolation
20 V
Line power
Line Voltage (selectable)
100 V/120 V or 220 V/240 V
Line Frequency
47 to 63 Hz
Line Voltage Variation
±
10 % about line voltage setting
Power consumption
15 VA maximum
Dimensions
Height 13.3 cm (5.25 in) plus 1.5 cm (0.6 in) four feet on bottom
Width
¾ standard rack width
Depth
47.3 cm (18.6 in) overall
Weight (without options)
4 kg (9 lb)