Table 1-4. Function Keys (cont)
No
Name
Description
D
SETUP
4
Press
J
to advance through the LCD Backlight, Interface, and
Address menus.
•
Use
M
and
O
to adjust LCD backlight when the LCD menu is
displayed.
•
Use
M
and
O
to toggle between Serial and GPIB interface when
the Interface menu is displayed.
•
Use
M
and
O
to scroll from Address:1 to Address 30 when the
address menu is displayed.
E
OUTPUT
7
Selects Output mode.
F
INPUT
8
Selects Input mode.
G
ZERO
9
Zeros the pressure module reading when in Pressure Measurement
mode.
Zeros the thermocouple TC mV/
°
C offset when in TC Measurement
mode.
H
°
C/
°
F
6
Toggles between Centigrade and Fahrenheit when you are using
the TC or RTD functions.
I
AUTOSET
3
AUTOSET runs through the setpoints you entered using the SET
function. Press
I
3
. AUTO SET POINT? appears on the
display.
Enter a number between 1 and 9 that corresponds to the number of
setpoints being used. DWELL TIME 5-500? appears on the display.
Dwell time is the number of seconds between each setpoint. The
output cycles through each setpoint and then reverses the order.
For example, if 5 is entered for the number of setpoints, the
Calibrator cycles through setpoints 1, 2, 3, 4, 5 and then reverses to
setpoints 4, 3, 2, and 1.
W
Caution
Setpoints of 30 V or greater will not go to standby
when you use this feature.
J
EXP
H
Used to enter an exponent when you define a custom RTD.