Remote Commands
Remote Command Listing
5
5-5
Error Code Listing (continued)
Error
Number
Message
Class
Description
118
CME
An invalid parameter was received.
120
EXE
The serial input buffer overflowed
121
EXE
The command string buffer overflowed.
122
QYE
The serial output buffer overflowed.
123
DDE
The output overloaded.
124
DDE
525A is out of tolerance. This error is set after a failed initialization or
a failed *TST?.
125
DDE
525A has a ADC failure. This error is set after a failed initialization or
a failed *TST?.
Remote Command Listing
The following is an alphabetical list of all Calibrator commands and queries,
including common commands and device-dependent commands. Each command
title includes a graphic that indicates remote interface applicability, IEEE-488 and
RS-232, and command group: Sequential and Overlapped.
IEEE-488 (GPIB) and RS-232 Applicability
IEEE-488
RS-232
x
x
Each command and query has a check box indicating applicability to IEEE-488
(general purpose interface bus, or GPIB) and RS-232 remote operations. For
sorting purposes, this list ignores the
*
character that precedes the common
commands.
Sequential Commands
Sequential
x
Commands executed immediately
as they are encountered in the data stream are called sequential commands. For
more information, see “Sequential Commands” in Chapter 4.
Overlapped Commands
Overlapped
x
Commands that require additional
time to execute are called overlapped commands because they can overlap the next
command before completing execution. To be sure an overlapped command is not
interrupted during execution, use the
*OPC
,
*OPC?
, and
*WAI
commands to
detect command completion. For more information, see “Overlapped Commands”
in Chapter 4.