(Local command) Puts the Calibrator into the local state, clearing the remote state
(see the
REMOTE
command) and front panel lockout (see the
LOCKOUT
command). This command duplicates setting the IEEE-488 REN line to false.
Parameter:
(None)
Example:
LOCAL
Set the instrument into the local state, clearing the remote state and front panel
lockout (if enabled).
LOCKOUT
Overlapped
IEEE-488
RS-232
Sequential
x
x
x
x
(Lockout command) Puts the Calibrator into the lockout state when in remote
control (see the
REMOTE
command). This means no local operation at the front
panel is allowed during remote control. To clear the lockout condition, use the
LOCAL
command. This command duplicates the IEEE-488 LLO (Local Lockout)
message.
Parameter:
(None)
Example:
LOCKOUT
Set the instrument into the front panel lockout state. The front panels controls
cannot be used.
*OPC
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Operations Complete command) Sets bit 0 (OPC) of the Event Status Register to
1 when all pending device operations are complete. Also see the
*ESR?
command.
Parameter:
(None)
Example:
*OPC
Set bit 0 of the Event Status Register to 1 when all pending device operations are
done.