Using The Bushealth Function
Input Connections and Tested Signals
6
75
Foundation H1 Bus
Default probe setting 10:1. Use the Fluke 10:1 probe.
1
Connect the red probe to test tool input A
2
Connect the probe ground lead to the H1 bus +
3
Connect the probe tip to the H1 bus -
DATA +
DATA -
Figure 41. Foundation H1 Bus Measurement
Connections
Table 7. Foundation H1 Bus Tested Signal Properties
Signal Description
V-Level Bias
Bias voltage level
V-Level Peak-peak
voltage
Data
Bit
width
Rise
Rise time as % of bit width
Fall
Fall time as % of bit width
Jitter Jitter
distortion
Signal Dist.
Signal distortion
Noise-HF
High Frequency noise >39.1 kHz
Noise
Mid Frequency noise 7.8 ….39.1 kHz
Noise-LF
Low Frequency noise < 7.8 kHz
Note
The bus normally has continuous data traffic.