Setting Trigger Level and Slope
The Connect-and-View™ function enables hands-off
triggering to display complex unknown signals.
When your test tool is in manual range, do the following:
Perform an auto set.
AUTO
appears
at the top right of the screen.
Automatic triggering assures a stable display of virtually
any signal.
From this point, you can take over the basic trigger
controls such as level, slope and delay. To optimize trigger
level and slope manually, do the following:
1
Display the
TRIGGER
key labels.
2
Trigger on either positive slope or
negative slope of the chosen
waveform.
Dual Slope Triggering ( X ):
19xC-2x5C versions can trigger on
both positive slope and negative
slope.
3
Enable the arrow keys for manual
trigger level adjustment.
Figure 27. Screen with all Trigger Information
4
Adjust the trigger level.
Observe the trigger icon that indicates the trigger
position, trigger level, and slope.
At the bottom of the screen the trigger parameters are
displayed (See
Error! Reference source not found.
) .
For example,
means that input A is used as
the trigger source with a positive slope.
When no trigger is found, the trigger parameters appear in
gray.
Содержание 19xC
Страница 2: ......
Страница 9: ...Contents continued v A Bushealth Measurements A 1...
Страница 10: ......
Страница 36: ...Fluke 19xC 2x5C Users Manual 26...
Страница 60: ...Fluke 19xC 2x5C Users Manual 50...
Страница 110: ...Fluke 19xC 2x5 Users Manual 100...
Страница 146: ...Appendices Appendix Title Page Bushealth Measurements A 1...
Страница 147: ......