Using The Bushealth Function
Input Connections and Tested Signals
6
71
AS-i Bus
Default probe setting 10:1. Use the Fluke 10:1 probe.
1
Connect the red probe to test tool input A
2
Connect the probe ground lead to the AS-i bus -
3
Connect the probe tip to the AS-i bus +
AS-i +
AS-i -
M12 Flat-C
1 Brown
Blue
3
Figure 37. AS-i Bus Measurement Connections
Table 3. AS-i Bus Tested Signal Properties
Signal Description
V-Level Bias
Bias voltage
V-Level Peak-Peak
voltage
Note
The bus normally has continuous data traffic.